Introduction
Acute Myocardial Infatction
BioTracer™ Cardiac 3 in 1 Rapid Test is intended for the qualitative determination of cTnI, CK-MB and Myoglobin in human serum, plasma or whole blood.
• Immunochromatography based one-step in vitro diagnostics test
• Point of care test at emergency room
• No instrument needed
• All material provided, Ready to use test
• No Cross reactivity
Troponin T
2,000ng/ml
Troponin C
1,000ng/ml
Human skeletal muscle Troponin I
300ng/ml
CK-MM
1,000ng/ml
CK-BB
1,000ng/ml
• Specimen : Serum , Plasma or Whole Blood
• Reading Time : 15~20 min
Troponin I
CK-MB
Myoglobin
Analytical sensitivity
1ng/ml
5ng/ml
70ng/ml
Sensitivity
96.8%
97.0%
98.7%
Specificity
97.5%
98.6%
97.7%
Test Procedure
• The myoglobin test results may be showed false negative in over 12 hours after onset of chest pain.
• A positive result will not change once it has been established at 15 minutes, however in order to prevent any incorrect results, the test results should not be interpreted after 20minutes.
Specification
Cat. No.
Product
Type
Pack Size
Specimen
Shelf Life
12011
Cardiac 3 in 1 (cTnI/CK-MB/Myo)
Card
10T/Kit
Serum/Plasma/Whole blood
24 M
Share Cardiac 3 in 1 in Social Media