Features
-Excellent image quality
-Compact design
-Easy maneuverability
-Economical solution
-Flexible application
-Short exposure time
-Wireless & wire remote exposure
Spec
PARTS
TECHNICAL SPECIFICATIONS
X-RAY GENERATOR
? High Frequency Inverter Control System
OUT PUT
? Adjustable Tube Voltage : 40kV~110kV (1kV step)
? mAs Range : 1.0mAs~200mAs
? Max Output Power : 5.2kW @ 60kV 86mA at Input 200V
X-RAY TUBE
? Focal Spot : 1.8mm
? Max. Tube Voltage / Current : More Than 110kV / 100mA
? Anode Type : Stationary Anode
? Anode Heat Storage Capacity : More Than 30kJ
? Target Angle : 15°
COLLIMATER
? Lamp Illumination : More Than 160 luk (24V / 150W)
? SID Scale : 2m
AC INPUT
? Voltage : AC 115 / 120 & 220 / 230V
? Phase & Frequency : Single - phase 50 / 60Hz
WEIGHT
? Net Weight : 150kg
Share MOBILE X-RAY SYSTEM in Social Media