Heart Plus ResQ AED (Automated External Defibrillator) for Sudden Cardiac Arrest
Cardiac arrest can happen to any one, at any place, at any time.
Why do we need AEDs?
Sudden Cardiac Arrest (SCA) is the leading cause of death in the United States and in several other countries claiming more than 350,000 lives each year in the U.S. alone. Approximately 92% of those who experience sudden cardiac arrest do not survive. It is unfortunate there are so many fatalities related to this condition in the world each year, but many victims can be saved with the timely and proper use of AEDs by first responders performing CPR (cardiopulmonary resusciation) until medical rescuers (EMS) arrive on scene.
Where should AEDs be installed?
- Recommended locations for AEDs
- Schools / Gyms / Public Facilities / Stadiums / Theatres / Medical Centres / Recreational Centers / Swimming Pools /
- Shopping Centres / Dental Office / Entertainment Venues / Sporting Venues / Pharmacies / Electrical Rooms / Tracks /
- Cruises / Airplanes / Subways / Bus Stations / Train Stations/ Religious Centres / Police & Emergency Service Vehicles and other places where many people congregate.
Why Heart+ResQ AED?
Heart+ResQ is a life-saving device that can be used safely and quickly in the case of an emergency.
- Our technology features accurate analysis to only advise shocks when necessary.
- If the electrodes are not attached properly, the electric shock process is halted.
- If the electric shock button isn’t pushed within 20 seconds, battery will discharge.
- Inclusion of an overcharge prevention circuit for proper electric shock loads.
- Dual electric shock waveform control circuit prevents malfunction of device.
- Anyone can use by following voice prompts : Voice prompts are simple and easy to follow.
- Easy to check the status of device: Check the status of device via status indicator, if some problem occurs.
- Simple management of device to ensure its readiness: Heart+ResQ performs a self diagnostics every day to check its vital components. Administrator pre-connects pads and closes the cover after each use.
- One pad for both adults and children: Save time by using same electrode pads on both adults and infants/children.
- Automatically analyzing accurate ECG: After attaching the pads to the patient, ECG is analyzed. An electric shock is advised if a shockable rhythm is detected.
Heart Plus ResQ Overview:
- Parts: Microphone, Status button, Pads port,Shock button, Internal speaker, Power button for adults, Power button for infants
Defibrillator Series: NT-381.C
- Set Configuration: Defibrillator, battery pack, sealed electrode, pads, Heart+ResQ User Guide, Protective Carrying Case, “NT-MPR” CD-ROM, Bluetooth, Adaptor (optional)
- Waveform: Truncated Exponential Biphasic
- Energy:150J (Adult, Impedance 50Ω), 50J (Pediatric, Impedance 50Ω)
- Shock Charge Time:9 seconds (±10%)
- Shock Method:Manual button (flashes when ready)
- Mode Conversion: Adult/Child toggle button, (Child Mode delivers 50J Pediatric shock for patients weighing less than 25kg,
- approximately 12 months to 8 years old)
Physical:
- Dimensions 8.0cm(H) x 22.0cm(W) x 28.2cm(D)
- Weight 2kg (including battery and pads)
Environmental:
- Water Resistance IP55 per IEC 60529
- Temperature Operating: 0°C to 40°C;
- Storage: -20°C to 60°C
- Humidity(operating/storage): 10% to 95% Relative Humidity, Non-Condensing
- Altitude (operating/storage): 0 to 5000 meters
- Drop Test:1.5 meters per IEC 60068-2-32
ECG:
- ECG Recording: LEAD II
- Analysis Overview: Patient evaluated to determine whether ECG rhythm represents shockable or non-shockable state. Shockable rhythms are ventricular fibrillation (VF) and ventricular tachycardia (VT).
- Analysis Time: Approx. 7 seconds (Read: about 5 seconds, Analyze: within 2 seconds)
Data:
- Stored Contents: ECG recordings, operation information (whether shock delivered or not), ambient noise, self-diagnostics history.
- Memory: ECG Records 5 sec during analysis only (up to 1000 instances), Ambient Noise: 60 minutes,4 inst of 15 min, Self-Test: Times & results of 3,000 events
- Storage Type: Internal Memory
- Access / Retrieval: "NT-MPR" App (for PC) via Bluetooth connection
Battery Check:
Battery Reserve Level:
- High : three (3) bars
- Medium : two (2) bars
- Low : one (1) bar; red Status LED
- Critical : no bars; red Status LED
Battery Status Audible Warnings:
- Low: Voice prompt states "Battery's
- voltage is low, check battery"
- Critical: Voice prompt states
- "Change cartridge"
Electrode Pads:
- Shelf Life: Two (2) years, for one-time use only
- Conductive Gel: PE Foam, Hydrogel
- Conductive Element: AgCl (Silver Chloride)
- Cable Length: 2.1 meters (±1%)
- Applicable Ages: All ages (12 months & older)
- Sealing: Airtight
- Size: 125mm x 95mm
- Pads Placement: Anterior-lateral (visual details on pads)
Battery Pack:
- Type: Non-rechargeable lithium (LiMn02)
- Capacity: Voltage DC 12V / 4,200mAH
- Standby Mode Life: Recommended five (5) years @ 25°C, 180 shocks @ 25°C, 130 shocks @ 0°C
- Dimensions: 2.2cm(H) x 16.4cm(W) x 5.4cm(D)
- Weight: 260g (±10%)
Self-Diagnostics Test:
-
Daily self-diagnostics tests ensure that circuits for measuring patient impedance, control circuit internal discharge of electric energy, and control circuit for charge/discharge of electric energy are functioning properly.
-
Upon completion of daily self-diagnostics test, results are automatically stored within device's internal memory.
Auditory/Visual Instructions:
- After device is powered on, step-by-step voice instructions guide first responder through resuscitation procedure, which includes pads placement, hands off times (analysis and shock process), and CPR instructions.
- CPR instruction prompts user to commence cardiopulmonary resuscitation, while flashing metronome light sets pace at 100 compressions per minute for 30 compressions with brief pause for 2 mouth-to-mouth breaths (5 repetitions), in accordance with American Heart Association 2010 guidelines.
Share Heart Plus ResQ AED (Automated External Defibrillator) in Social Media